PET INTERSPACED ANTI-SCATTER GRIDS

PET interspaced anti-scatter x-ray grids (hereafter 'poly grid') was developed through more than 40 years experience of manufacturing the anti-scatter grids.
Our poly grid provides you with better images with dramatically reduced x-ray doses due to PET interspacer and a new manufacturing technology.
Available specifications
- Size / Shape : all sizes & shapes
- Lines/cm : 40 lines/cm - 80 lines/cm
- Ratio : 5 : 1 - 15 : 1
- SID : 60cm - infinite
- Skin material : Carbon / Aluminum
Advantages
■ Dramatically low dose
- Dose is dramatically lower than aluminum interspaced grids.
■ Super primary transmission
- PET interspacer allows around 30% better in primary x-ray transmission.
■ Less weight
- By using PET as interspacer of the grid, our poly grid is 20-30% less weight.
■ Excellent uniformity with no moire effect
- Lines straightness is secured due to a new manufacturing technology.
■ Design customization
- Poly grid can be customized in design.

( Design customization )

( without cover / with carbon cover )
Physical Characteristics
Contrast Improvement
(K)
Properties
Primary Transmission
(Tp)
Bucky Factor = Dose
(B)
PET interspaced grid
( 70 lines / cm, r10 )
72
2.18
3.03
PET interspaced grid
( 43 lines / cm, r10 )
84
2.70
2.24
-
The test was conducted under IEC60627 2001 (second edition).
-
Test results may vary depending on the test environment.
-
This test results are presented for reference purpose only.
Applications

C-ARM
IN OPERATION ROOM
FOR DOSE REDUCTION

FLUOROSCOPY
FOR DOSE REDUCTION

GENERAL X-RAY
FOR IMAGE OPTIMIZATION